Adjusting SPAN on Model 615

Adjusting SPAN on Model 615

Adjusting SPAN on Model 615


This is method can only be used if calibration is already within 10%.  For larger adjustments see - https://support.scale-tec.com/portal/en/kb/articles/calibrating-a-model-615-indicator-using-a-custom-sensitivity
1.      With the scale empty and zeroed out, press the hidden programming key for 5 seconds until "615" is displayed.

 

 Some Model 615's may have NET/GROSS buttons that replace the hidden arrow keys.


2.      Press SETUP repeatedly until "SPAN" is displayed.
3.      Press PRINT/SELECT.  The scale should show 0.
4.      Place a known weight on the scale.  A heavier known weight will result in a more accurate calibration.
5.      Use the UP ARROW and the LEFT ARROW to increase or decrease the displayed value to match the known weight on the scale.
6.      Once the correct weight is displayed, press PRINT/SELECT followed by ZERO to return to gross weigh mode.

    • Related Articles

    • Using Span to Fine Tune the Calibration on a 640

      Using Span to Fine Tune the Calibration on a 640 Span should only be used for relatively small calibration adjustments of 10% or less Press and hold the "Hold/Menu" button for three beeps then release. Enter "640".  Use "RM" to increase the displayed ...
    • Entering a Configuration Number on a Model 715 Indicator

      Entering a Configuration Number on a Model 715 Indicator This is the best method to calibrate a 715 if using AWT loads cells of a known calibration diameter. 1.      Type in 715 2.      Hold the hidden E key for two seconds until 715E is displayed. ...
    • Calibrating a Model 615 Indicator Using a Custom Sensitivity

      Calibrating a Model 615 Indicator using a Custom Sensitivity This is the best method to calibrate a 615 if using any loads cells with a known output. 1.      Press the hidden programming key for 5 seconds until "615" is displayed.    Some Model 615's ...
    • Entering a Configuration Number on a Model 615 Indicator

      Entering a Configuration Number on a Model 615 Indicator This is the best method to calibrate a 615 if using AWT loads cells of a known calibration diameter. 1.      Press the hidden programming key for 5 seconds until "615" is displayed.    Some ...
    • Entering a Configuration Number on a Model 915 Indicator

      Entering a Configuration Number on a Model 915 Indicator This is the best method to calibrate a 915 if using AWT loads cells of a known calibration diameter. 1.      Type in 915 2.      Hold the hidden E key for two seconds until 915E is displayed. ...