EZ150, EZ210 and EZ350 Setup and Calibration

EZ150, EZ210 and EZ350 Setup and Calibration


    • Related Articles

    • GT400 and UNVERFERTH 410 Calibration

      Trade up to a POINT and get $500 value for your old indicator! Click here now for more information Qualification: Any scale brand and scale model qualifies. Any condition qualifies. (Yes even broken scale indicators) Any age of scale indicator ...
    • Changing Setup and Calibration Numbers in a Digi-Star Indicator

      Trade up to a POINT and get $500 value for your old indicator! Click here now for more information Qualification: Any scale brand and scale model qualifies. Any condition qualifies. (Yes even broken scale indicators) Any age of scale indicator ...
    • Avery Weigh-Tronix 640 Indicator Calibration

      Trade up to a POINT and get $500 value for your old indicator! Click here now for more information Qualification: Any scale brand and scale model qualifies. Any condition qualifies. (Yes even broken scale indicators) Any age of scale indicator ...
    • Avery Weigh-Tronix Load Cell Calibration Numbers

      The reading @ 0.4mV/V can be used as the Calibration number for a Digi-star indicator, or the Custom number on a Weigh-tronix indicator.
    • Calibration Number Equation

      Here is the equation to figure out what the calibration number would be. .4 mv/v    X    Capacity    X    # of Bars    /    mv/v Output of the weigh bar  =    good starting Calibration # Example: We will be using a 141907 weigh bar in this example.  ...