EZ150, EZ210 and EZ350 Setup and Calibration

EZ150, EZ210 and EZ350 Setup and Calibration


    • Related Articles

    • GT400 and UNVERFERTH 410 Calibration

      GT400 and Unverferth 410 Calibration Video
    • Changing Setup and Calibration Numbers in a Digi-Star Indicator

      Changing Setup and Calibration Numbers in a Digi-Star Indicator 1. Hold the ZERO button and the ON button until "SETUP" is shown on the screen. 2. After "SETUP" the indicator will display its current setup number with the last digit flashing. A good ...
    • Avery Weigh-Tronix 640 Indicator Calibration

      Watch this video to calibrate your indicator. Press and hold the "Hold/Menu" button for three beeps then release. Enter "640". Use "RM" to increase the displayed value and "M+" to decrease it. Press the "Hold/Menu" button to move one digit to the ...
    • Avery Weigh-Tronix Load Cell Calibration Numbers

      The reading @ 0.4mV/V can be used as the Calibration number for a Digi-star indicator, or the Custom number on a Weigh-tronix indicator.
    • Calibration Number Equation

      Here is the equation to figure out what the calibration number would be. .4 mv/v    X    Capacity    X    # of Bars    /    mv/v Output of the weigh bar  =    good starting Calibration # Example: We will be using a 141907 weigh bar in this example.  ...